Every behaviour of the simulated column follows the electron optics that Ruska described in 1934. The image is computed from these rules each frame.
LensA coil is a lens
In 1926 Hans Busch showed that the magnetic field of a short coil bends electrons the way a glass lens bends light. Its focal length follows f ∝ U / (N I)²: turn up the current and the lens gets stronger. Ruska wrapped the coil in iron with a narrow gap, the pole piece, squeezing the field into a few millimetres. His objective focused 75 kV electrons at about 6 mm.
WaveWaves 100,000 times shorter than light
Ruska learned of de Broglie’s matter waves in 1931 and at first feared a new wavelength limit. The arithmetic reassured him: at 75 kV, λ = h / √(2meU) ≈ 4.3 pm, five orders of magnitude below visible light. In 1932 Knoll and Ruska predicted a limit of 2.2 Å. It was reached about 40 years later.
FocusChange the voltage, lose the focus
Because f depends on the voltage, any wander of the high voltage moves the focus, and its ripple smears every point into a disc (chromatic aberration). The gas discharge had to burn "extremely steadily". Watch the HV ripple bar grow when you open the discharge.
RotationThe image turns as it grows
Electrons spiral through a magnetic lens, so the image rotates by an angle that grows with the coil current. Turn the projector coil and the specimen spins as it magnifies. Ruska’s operators lived with it; modern microscopes compensate for it.
Light budgetMagnification costs light
The image spreads the electrons over an area M² times larger. At 10,000× the specimen receives 108 times the current density of the screen. To see anything, the condenser must squeeze the beam onto a few micrometres. That heats the specimen, and an organic one chars.
ContrastContrast by scattering, not absorption
Ruska noticed that even very thin specimens gave contrast. Not by absorbing electrons, but by scattering them out of the lens aperture. Thin specimens absorb little and stay cooler, which made biological electron microscopy possible.